Philips Respironics V60 Ventilator

  • The Philips Respironics V60 Ventilator is the gold standard hospital dedicated noninvasive ventilator. With advanced breath delivery technology and algorithms designed to maximize performance in a leak prone environment, V60 adapts to your patients’ ventilation needs.

    • Advanced NIV with Auto-Trak: Provides auto-adaptive leak compensation, inspiratory triggering, and expiratory cycling to deliver optimal synchrony in the face of constantly changing leaks and patient demand-alleviating constant bedside supervision and manual adjustments.
    • Pre-defined mask settings save time: The Philips Respironics V60 supports the success of NIV by increasing the speed and ease of treatment initiation. Its pre-defined settings for Philips Respironics hospital masks help you save time. You select the setting, and the V60 automatically calibrates flow characteristics for better monitoring and therapy.
    • CPAP with C-Flex for greater patient comfort: CPAP (Continuous Positive Airway Pressure) with C-flex offers three levels of flow-based expiratory pressure relief. This option leads to improved sleep quality and patient comfort, adding greater flexibility and improved treatment acceptance.
    • Multiple treatment modes for a wide range of patients: The Philips Respironics V60 helps you meet the needs of a broader patient population. Designed to include pediatric use and equipped with several modes, the V60 allows you to meet the specific needs of your patients.
    • Expansive colour touchscreen for easy operation: The Philips Respironics V60 has a high-resolution color touchscreen. The high level of visual detail makes operation easy and facilitates waveform interpretation.
    • PCV increases control: PCV can be used when greater control of the patient's breathing pattern is required.
    • AVAPS: AVAPS (Average Volume Assured Pressure Support) maintains a target tidal volume in a pressure limited mode. It provides extra assurance similar to a volume limited mode with the safety of a pressure limited mode.
    • Standby mode supports patient/clinician interaction: Focus on your patients without being interrupted by nuisance alarms by switching the V60 to standby mode.
    • PPV for variable breathing patterns: Proportional Pressure Ventilation (PPV) provides inspiratory flow and pressure in proportion to the patient's spontaneous effort*, improving patient control over their ventilation**, and improving comfort.
    • Transport flexibility: Internal 6-hour battery enables intra-hospital transport
    • Auto-Trak Plus for custom treatment criteria: Auto-Trak Plus, an option available in the S/T, PC, AVAPS, and PPV modes, is for the subset of patients who may benefit from a customized titration of triggering and cycling criteria. In certain patients with extremely weak efforts or in others with low compliance, Auto-Trak Plus lets clinicians make finely-tuned adjustments to the base Auto-Trak algorithm to achieve patient ventilator synchrony.
    • Enhanced communications: Data communication capabilities support hospital connectivity
    • Respi-Link* for efficient upgrades: Through this remote connection tool, you can perform efficient system diagnostics and easily apply upgrades via internet.
  • Dimensions

    Dimensions listed below with the following configuration: V60 Ventilator with backup battery, ventilator stand, O2 tank holder, circuit arm with mount, patient circuit, O2 manifold with hoses.

    • Height: 53.5” (136 cm)

    • Width: 25” (64 cm)

    • Weight: 79.5 lbs (36.1 kg)

    Dimensions listed below for only the V60 Unit.

    • Height: 13.3” (33.7 cm)

    • Width: 15.5” (39.4 cm)

    • Depth: 16.5” (42.9 cm)

    • Weight (With optional battery): 25.7 lbs (11.7 kg)

    • Weight (Without battery): 23.3 lbs (10.6 kg)

    Patient Types

    • Adult

    • Pediatric
      •(Less than or equal to 20 kg)

    Battery

    • 14.4V, 10.0 Ah, 163 Wh

    • Maximum system current draw: 11 A

    • Charge Voltage: 16.9 V Maximum

    • Operating Time: 360 minutes under normal condition

    Modes

    • Continuous positive airway pressure (CPAP)

    • Spontaneous with timed backup (S/T)

    • Pressure control ventilation (PCV)

    • Average volume assured pressure support (AVAPS) —optional

    Settings

    • C-Flex OFF, 1–3

    • CPAP 4–25 cmH2O

    • EPAP 4–25 cmH2O

    • IPAP 4–40 cmH2O

    • I-time (inspiratory time) 0.30–3.00 sec

    • Max P (AVAPS maximum IPAP) 6–40 cmH2O

    • Min P (AVAPS minimum IPAP) 5–30 cmH2O

    • O2 (oxygen percent) 21–100%

    • Ramp time OFF, 5–45 min

    • Rate (respiratory rate) 4–60 BPM

    • Rise (rise time) 1–5

    • Triggering and cycling Auto-adaptive (Auto-Trak)

    • AVAPS target tidal volume 200–2000 ml

    Monitored Parameters

    • Breath Phase / Trigger Indicator Spont, Timed, Exhale

    • PIP 0–50 cmH2O

    • Patient / Total Leak 0-200 l/min BTPS

    • Patient trigger 0–100%

    • Respiratory rate 0–90 BPM

    • Ti/Ttot 0–91%

    • Minute volume ²OPLQ%736

    • Tidal volume 0–3000 ml BTPS

    General

    • Oxygen inlet pressure range 276–600 kPa (40–87 psig)

    • Weight 10.9 kg (24lb) with optional battery; 10.0 kg (22 lb) without optional battery

    • Dimensions 33.7 cm (13.3 in) height; 39.4cm (15.5 in) weight; 42.9 cm (16.5 in) depth

    Electrical

    • AC voltage 100–240 VAC

    • AC frequency 50–60 Hz

    • AC power 300 VA

    • Battery operating time 6 hours in normal conditions

    Environmental Specifications (Operating)

    •  Temperature: 41 to 104 ºF (5 to 40 ºC)

    • Relative Humidity: 15 to 95% (noncondensing) Barometric Pressure: 600 to 765 mmHg (approximately -61 to 1951 m (-200 to 6400 ft) relative to sea level) 

    Environmental Specifications (Storage)

    • Temperature: -4 to 122 ºF (-20 to 50 ºC)

    • Relative Humidity: 10 to 95% relative (noncondensing) Barometric Pressure: 600 to 765 mmHg (approximately -61 to 1951 m (-200 to 6400 ft) relative to sea level) 

    High-pressure Oxygen Supply

    • Connector: DISS male, DISS female, NIST

    • Pressure: 2.76 to 6.00 bar / 276 to 600 kPa / 40 to 87 psig

    • Flow: 175 SLPM

    • Connector: SIS

    • Pressure: 3.31 to 6.00 bar / 331 to 600 kPa / 48 to 87 psig

    • Flow: 175 SLPM

    High-pressure Oxygen Supply (Using V60 manifold)

    • Connector: DISS male, DISS female, NIST

    • Pressure: 3.10 to 6.00 bar / 310 to 600 kPa / 45 to 87 psig

    • Flow: 175 SLPM

    • Connector: SIS

    • Pressure: 3.66 to 6.00 bar / 366 to 600 kPa / 53 to 87 psig

    • Flow: 175 SLPM

    Inspiratory Outlet (To patient port)

    • Connector: ISO 15 mm female/22 mm male conical

    Air Supply

    • Integrated blower

    Flows and Pressures

    •    Flow Delivery: 150 L/min at 40 cmH2O at 1951 m (6400 ft) altitude (10% degradation in flow at 2286 m (7500 ft))

    • Flow Range: -240 to 240 L/min BTPS

    • Pressure Range: 4 to 40 cmH2O

    • Dynamic pressure regulation: ± (2 cmH2O + 4% of target)

     